Voltage Probe Manual and Data

North Star Research Corporation

 

Standard Voltage Probe Data

High Voltage Probes Up to 400 kV

 

High Voltage Probes

High voltage measurements have a variety of different requirements, and our PVM series and VD series probes are designed for most of these requirements. The PVM probes are designed for wide bandwidth and ease of portability where a variety of measurements must be made. The VD series probes are designed for use at higher voltage ranges, and where physical stability is required.

Specialty Probes PVM series probes with PVM-5 on top and PVM-10 on the bottom

The design of the probes is unique. The probe comes fully adjusted without any requirement to further adjust the probe in the field. Stray capacitances internal to the probe are factory calibrated. Stray capacitances due to proximity effects are minimied by the unique conductor placement.

Low temperature and voltage coefficient resistors and capacitors are used throughout the probe to provide maximum measurement accuracy. The probe has a rugged nylon housing, and it is filled with dielectric oil for maximum high voltage stability.

Our PVM Series probes our available with either 1000:1 or 2000:1 calibration factors at the time of order.

Voltage Probe Specifications

PVM Series Portable High Voltage Probes to 60 kV DC

PVM series high voltage probes are designed for general use, and for exceptional high frequency response. The probes have applications ranging from automotive ignition to excimer laser system measurement to EMI measurement. They are factory calibrated, and they do not require adjustment. In general the probes are for use with 1 Megohm oscilloscopes, but we also offer an optional switch which can compensate for various measurement instruments such as 10 Megohm meters as well. These units are intended for a wide range of applications where portability and ease of use are essential.

 

Model Number

PVM-1

PVM-2

PVM-3

PVM-4

PVM-5

PVM-6

PVM-11 (PVM-10)

PVM-12

Max DC/Pulsed V (kV)

40/60

40/60

40/60

40/60

60/100

60/100

10/12

25/30

Max Frequency (Mh.)

80

80

40

110

80

80

50

80

Cable Impedance (ohms)

50

50

50

93

50

50

50

50

DC - 2 H.accuracy

<0.1%

<0.1%

<0.1%

<0.1%

<0.1%

<0.1%

<0.1%

<0.1%

2 H. - 200 H. accuracy

<1 %

<1. %

<2. %

<1. %

<1 %

<1 %

<1.5 %

<1.5 %

200 H. - 5 Mh. accuracy

<1.5%

<1.5%

<3%

<1.5%

<1.5%

<1.5%

<2.%

<2.%

> 5 Mh. Accuracy

<3%

<3%

<4%

<5%

<3%

<4%

<4%

<4%

Input R/C (Megohm/pf)

400/13

400/13

400/10

400/10

400/12

400/12

100(50)/15

300/7

Cable Length (ft./m)

15/4.5

30/9

100/30

15/4.5

15/4.5

30/9

15/4.5

15/4.5

Standard Divider Ratio

1000:1

1000:1

10,000:1

1000:1

1000:1

1000:1

1000:1 (100:1)

1000:1

Length (inches/cm.)

19/47

19/47

19/47

19/47

19/47

19/47

7/18

9/23

*美NORTH STAR 高壓探頭

Add -2 to any PVM-1 - PVM-6 or PVM-12 part number for 2000:1 ratio

VD Series High Voltage Probes 60 to 300 kV DC

Specialty Probes

VD series high voltage probes are floor standing high voltage probes which are designed for rugged day in - day out use. They are used in a wide range of applications ranging from evision tube manufacturing to radar to advanced particle accelerator applications. Resistors with an extremely low voltage coefficient of resistance are used, and all capacitors are temperature, frequency, and voltage stabilied for the best possible performance. The probes all have field defining toroids as a standard item in order to minimie the proximity effect (stray *美NORTH STAR 高壓探頭capacitance) and maximie the reproducibility of the measurement. The high and low frequency calibrations are carefully matched before shipment. Very high frequency cable effects are also carefully compensated so accurate measurements can be made even when the cable length exceeds the pulse duration. No adjustments are necessary once the probes have been factory calibrated.

Model Number

VD-60

VD-100

VD-150

VD-200

VD-300

VD-400

Max DC/Pulsed V (kV)

60/120

100/180

150/240

200/300

300/420

400/550

Max Frequency (Mh.)

20

20

20

16

12

8

Cable Length (ft.)

30

30

30

30

30

30

DC accuracy

<0.1%

<0.1%

<0.1%

<0.1%

<0.1%

<0.1%

10 H. - 1 Mh. Accuracy

1 %

1 %

1 %

2 %

2.5 %

3 %

>1 Mh Accuracy

3 %

3 %

3 %

3 %

4 %

4 %

Resistance (Megohms)

800

1600

2000

2800

4500

7000

Height (inches/cm.)

20/50

24/60

30/75

40/99

54/135

72/180

Diameter (in/cm.)

11/28

11/28

12/29

16/40

24/61

24/61

Capacitance (approx. pf)

27

25

27

24

20

16

Base Diameter(in/cm.)

10/25

10/25

12/30

20/50

30/76*

30/76

Standard Divider Ratio

10,000:1

10,000:1

10,000:1

10,000:1

10,000:1

10,000:1

 

留言框

  • 產(chǎn)品:

  • 您的單位:

  • 您的姓名:

  • 聯(lián)系電話:

  • 常用郵箱:

  • 省份:

  • 詳細(xì)地址:

  • 補(bǔ)充說明:

  • 驗(yàn)證碼:

    請(qǐng)輸入計(jì)算結(jié)果(填寫阿拉伯?dāng)?shù)字),如:三加四=7
北京志鴻恒拓科技有限公司

北京志鴻恒拓科技有限公司

地址:北京市朝陽區(qū)建國路88號(hào)soho現(xiàn)代城

© 2025 版權(quán)所有:北京志鴻恒拓科技有限公司  備案號(hào):京ICP備16029562號(hào)-5  總訪問量:589090  sitemap.xml  技術(shù)支持:化工儀器網(wǎng)  管理登陸

浓毛老太bbwbbwbbwbbw看| 日韩人妻无码中文字幕视频| 波多野结衣网站| 国产精品久久久久久一区二区三区| 国产AV国片精品一区二区| 性欧美残忍折磨猛交无码视频| 边啃奶头边躁狠狠躁小说| 男女xxoo娇喘呻吟在线观看| 一个妈妈的女儿bd在线观看| 国产成人久久av免费| 55岁大妈玩csgo| 无码人妻丰满熟妇啪啪欧美| 张柏芝阿娇全套无删减1313| h高潮嗯啊娇喘抽搐视频a片小说| 亚洲AV无码国产精品色午夜软件| 麻豆人妻少妇精品无码区| 一性一交一伦一片a片庆| 亚洲精品无码久久一线| 不戴套干张雅丹| 又湿又黄裸乳漫画无遮挡网站| 无码少妇一区二区三区芒果| 亚洲欧美中文日韩在线v日本| 国产男女无遮挡猛进猛出| 国精品无码一区二区三区在线| 小sb几天没做sao死了| 特级做a爰片毛片免费看| 精国产品一区二区三区a片| 男生和女生一起打扑克| 男按摩师按摩奶头是正常的吗| 扒开老师内衣吸她奶头动态图| 超薄丝袜足j好爽在线| 一层一层剥掉女生的衣服的游戏| 欧洲人激情毛片无码视频| 无码精品av久久久奶水| 国内精品久久久久久久小说| 再深点灬舒服灬太大了添a片v| 扒开老师内衣吸她奶头动态图| 18禁白丝喷水视频www视频| 性无码专区无码| 最近中文字幕大全在线看| 午夜福利在线观看午夜电影街bt|